1. Open the Buffer Compartment drawer. Select the forward arrow to proceed.

2. Carefully remove the waste reservoir and open the waste port without spilling. Be aware reservoir may be full.

WARNING
Make sure to always wear gloves when handling waste.

3. Empty the contents into the appropriate hazardous waste container.

4. Reinstall the waste cap.

5. Using a dry low-particulate wipe, clean any spills on the waste reservoir and any liquid in the drawer.

6. Replace the empty waste reservoir in the Buffer Compartment.

7. Change gloves.

8. Select the forward arrow to proceed.

WARNING
This set of reagents contains formamide, a suspected carcinogen and reproductive toxin, and other potentially hazardous chemicals. Personal injury can occur through inhalation, ingestion, skin contact, and eye contact. Wear protective equipment, including eye protection, gloves, and laboratory coat appropriate for risk of exposure. Handle used reagents as chemical waste and discard in accordance with applicable regional, national, and local laws and regulations through a licensed hazardous waste contractor.