Create one sample sheet for each flow cell. There are two ways to generate a sample sheet:

Use the Singular Genomics online Sample Sheet Generator (recommended).

Generate a sample sheet manually as described below.

NOTE

You do not have to upload a sample sheet. You can also define the samples manually when you set up the run on the ICS, or not provide sample data at all. However, without index information the Instrument does not perform on-instrument demultiplexing.

1. Download a sample sheet template from the Instrument, C:\Program Files (x86)\Singular Genomics\G4 Instrument Control Software\RunTemplates.

NOTE

If you do not plan on specifying demultiplexing options, use the template without Demux section (FlowCellTemplate.csv). Do not use FlowCellTemplateDemuxFields.csv, since you need to edit the demultiplexing options in order for that template to work properly.

2. Fill out the following information for the Header section. Fields marked with an asterisk * are required.

Field

Description

Date

Date of sample sheet edit.*

Run Name
Run name.*

User Name

Name of operator.

User Email

Email of operator.

Workflow

Allows you to track the analysis pipeline.

Assay

Allows you to track the library prep used.

Run Notes

Any other notes for the run.

3. Fill out the following information for the Settings section. Fields marked with an asterisk * are required.

Note

Do not enter an index length that is longer than the physical length of the indices. This may interfere with the binding of the sequencing primers in subsequent sequencing steps.

Description

[read 1]

Length of read 1 in bp, maximum 150 bp.*

[read 2]

Length of read 2 in bp, maximum 150 bp.

[Index 1]

Length of Index 1 in bp, minimum 8 bp, maximum 24 bp.

[Index 2]

Length of Index 2 in bp, minimum 8 bp, maximum 24 bp.

Custom Primer Read1 (Y)

If a custom sequencing primer is needed for read 1 enter Y. If not, leave blank.

Custom Primer Read2 (Y)

If a custom sequencing primer is needed for read 2 enter Y. If not, leave blank.

Custom Primer Index1 (Y)

If a custom sequencing primer is needed for Index 1 enter Y. If not, leave blank.

Custom Primer Index2 (Y)

If a custom sequencing primer is needed for Index 2 enter Y. If not, leave blank.

Note

Custom sequencing primers can only be set at the run level, you cannot use custom sequencing primers on one flow cell and Singular Genomics sequencing primers on another flow cell within the same run. Note also that different lanes within a flow cell cannot have different sequencing primers.

4. Fill out the following information for the samples section. Note that within a single lane, no two samples can have the same combination of Sample_ID, Index1_Name, Index2_Name, Index1_Sequence and Index2_Sequence. Fields marked with an asterisk * are required.

Description

Sample_ID

Sample ID.*

Index1_Name

Name of Index 1. Leave empty if no Index 1 is used.

Index1_Sequence

Sequence of Index 1. Required for demultiplexing. Leave empty if no Index 1 is used.

Index2_Name

Name of Index 2. Leave empty if no Index 2 is used

Index2_Sequence

Sequence of Index 2. Required for demultiplexing. Leave empty if no Index 2 is used

Lane

Lane number, a single number from 1 to 4. Required.

Lane_Name

Lane name.

Project

Allows you to organize libraries in a project.

Loading_Concentration

Loading concentration on lane.

Application

Allows you to track the library prep used.

Notes

Any other notes for the sample library.

Reference

Library reference.

Note

If you want to sequence a library in multiple lanes, copy the line for that library in the sample sheet and change the lane number.

5. Name the sample sheet and save as CSV file (*.csv) in the SampleData folder or in a location available on the same network as the instrument. If you save the sample sheet on a network location, you may need to specify the permissions required for the folder on the instrument.

Note

For custom demultiplexing options, add a Demux section in between the Settings and Data sections and specify options as described here: How Do I Specify Custom Demultiplexing in the Sample Sheet?.